6-12 VS 9/VS 8 Vital Signs Monitor Service Manual
Pull the connection belt of the battery door at the position shown in the figure on the
rear cover to replace the battery door.
Uncover the tape that fixes the connection cable at the position shown in the figure on
the rear cover to replace the connection cable (6012) (power button connection cable)
between the main board and the alarm lamp board. After replacement, the tape needs to
be re-attached to fix the cable.