6-36 VS 9/VS 8 Vital Signs Monitor Service Manual
6.4.4
(D) Detaching the Main Frame and Front Cover Assembly
1. Unplug the sockets connecting the Wi-Fi antenna, Bluetooth antenna, Bluetooth
connection cable, SpO
2
cable, CO
2
cable, and power connection cable to the main
board. If the monitor is equipped with a multi-function connector, remove the
connector board connection cable socket indicated by the arrow on the main board.
Otherwise, the cable does not exist and you can skip the step.
2. Unscrew the M4*8 combination screw 1, unplug the pump/valve cable socket 2, and
unplug the two air tubes 3 connecting to the sensor on the main board.