Chapter 3 Product Specifications
27
3.3 Power Source Specifications
AC mains
Input voltage
100 to 240V
Frequency 50/60Hz
Power 140VA
DC mains
Input voltage
10 to 16VDC
(
rating-12 VDC
)
Power 80W
Internal battery
Number of batteries
1
Type
Sealed lead-acid battery or lithium-ion battery
Time to shutdown
5 to 15min (after the first low-power alarm)
Sealed lead-acid battery
Nominal voltage
12VDC
Capacity 2.3Ah
Operating time
75 minutes typical when powered by one new fully-charged
battery (25
℃
, ECG, SpO2, NIBP measurement per 15 minutes).
Charge time
About 8 h (in the running status or standby mode)
Lithium battery
Rated voltage
11.1VDC
Capacity
4.4Ah
Operating time
150 minutes typical when powered by one new fully-charged
battery (25
℃
, ECG, SpO2, NIBP measurement per 15 minutes).
Charge time
About 8 h (in the running status or standby mode)