Chapter 3 Product Specifications
31
3.7
Signal Output Specifications
Standards
Meets the requirements of EC60601-1 for short-circuit
protection and leakage current
Output impedance
50
Ω
ECG analog output
Diagnostic mode:
0.05 to 100Hz(12-lead: 0.05 to
150Hz)
Monitor mode:
0.5 to 40Hz
Bandwidth
(-3dB; reference
frequency: 10Hz)
Surgery mode:
1 to 20Hz
Signal delay
≤
25ms
Maximum propagation delay
25ms (In DIAGNOSTIC mode, NOTCH is OFF)
Sensitivity 1V/mV±5%
PACE rejection/enhancement No pace rejection or enhancement
IBP analog output
Bandwidth
0 to 12.5 Hz (-3dB, reference frequency: 1Hz)
Maximum propagation Delay
55ms (the filter function is disabled)
Sensitivity 1
V/100mmHg±5%
Nurse call output
Driver Relay
Electrical specifications
≤
60W,
≤
2A,
≤
36VDC,
≤
25VAC
Isolation voltage
> 1500 VAC
Signal type
Normally open or normally closed, selectable
Defibrillator synchronization pulse
Maximum time delay
35ms (R-wave peak to leading edge of the pulse)
Amplitude
3.5 V (min) at 3 mA sourcing; 0.8 V (max) at 1 mA sinking
Pulse width
100 ms ±10%
Rising and falling time
< 3ms
VGA
Signal
RGB: 0.7 Vp-p/75
Ω
;
Horizontal/vertical synchronization: TTL level