Chapter 5 Test and Material List
57
calibration, and then exit the IBP PRESSURE CALIBRATE menu.
Set the BP value of the simulator respectively to 40mmHg, 100mmHg and
200mmHg. Then the screen of the PM-7000 should display 40±1mmHg,
100±2mmHg and 200±4mmHg.
Set the simulator output to ART wave. Then the screen of the PM-7000 should
display relevant waveform properly.
Unplug the IBP probe. Then the screen should prompt “IBP: Transducer 1 OFF!”
and “IBP: Transducer 2 OFF!”
Plug the OHMEDA cable to the IBP1 channel. Then the prompting message “IBP:
Transducer 1 OFF!” disappears.
IBP2 test:
②
Plug the IBP cable to the IBP2 channel, and repeat the procedure in Section .
①
5.1.8 CO
1. Test fixture
Physiological signal simulator
2. Test procedure
Injectate and blood temperature test: Assemble the TB and TI test fixture, output
three TB temperature values: 36 , 37 and 38 . Then TB should be respectively
℃
℃
℃
36.0±0.1 , 37.0±
℃
0.1 and 38.0±0.1 . Set the injectate switch to ON, output two
℃
℃
TI temperature values: 0 and 2 . Then the screen should display 0±0.1 and
℃
℃
℃
2.0±0.1 .
℃
CO measurement: Set the CO.CONST and T
I
to the default values: 0.542 and 0 ,
℃
set the injectate switch to OFF, and then press START. Then the simulator will
output 0 , 2.5L/M and 0 , 5L/M within 2s. The CO values should be 2.5±0.25L/M
℃
℃
and 5±0.5L/M.
5.1.9 CO
2
1. Test fixture
CO
2
steel bottle (containing 10% CO
2
)
2. Test procedure
Mainstream CO
①
2
measurement: Set the calculation compensation of PM-7000 to
COMMON.
Plug the mainstream transducer to the CO
2
socket, connect the windpipe connector
with the CO
2
steel bottle, and open/close the valve of the CO
2
steel bottle based on
the interval of 3s. The CO
2
value should be the calibration gas pressure value:
76±5%mmHg. When the valve is opened permanently, the patient monitor prompts
“APNEA ALARM”.
Unplug the transducer. The patient monitor prompts “CO
2
transducer OFF” on
the main screen. Plug the transducer again. The patient monitor prompts “CO
2
transducer pre-heated”.
Sidestream CO
②
2
measurement: Set the calculation compensation of PM-7000 to
COMMON.
Plug the water trap to the water trap socket, connect the sampling tube with the CO
2
steel bottle, and open//close the valve of the CO
2
steel bottle based on the interval of
3s. The CO
2
value should be the calibration gas pressure value: 76±5%mmHg. When