Chapter 6 Maintenance and Cleaning
67
Chapter 6 Maintenance and Cleaning
6.1 Maintenance
6.1.1Checking Before Using
■
Check the patient monitor for mechanical damages;
■
Check all exposed conductors, connectors and accessories;
■
Check all functions that are possibly enabled for the monitored patient, and
ensure the device is in good working status.
In case of any damage, stop using this patient monitor, and contact biomedical
engineers of the hospital or Mindray maintenance engineers.
6.1.2 Regular Checking
An all-around check, including the safety check, should be done by qualified personnel
every 6-12 months or after maintenance each time.
All checks in which the patient monitor should be disassembled should be done by
qualified maintenance personnel. The safety and maintenance checks can be done by
Mindray engineers. The local office of Mindray at your region will be pleased to provide
you with the information about the maintenance contract.
6.2
Cleaning
Do switch off the patient monitor and disconnect the AC power supply before
cleaning it or the probes.
The PM-7000r should be dust free. To clean the surface of its enclosure and screen,
use the cleaning agent that is not corrosive, for example, soap and water.
1. Do not use strong solvent, such as acetone;
2. Most cleaning agents must be diluted before being used, so conduct dilution under
the instruction of manufacturers;
3. Do not use any erosive material (such as steel wool or polishing agent);
4. Prevent the ingress of any liquid to the enclosure and any part of the device;
5. Ensure no residue of cleaning liquid on the surface of the device.
6.3
Cleaning Reagent
1. Diluted aqua ammonia
2. Diluted sodium hypochlorite (bleaching powder for washing)
3. Hydrogen peroxide 3%
4. Ethanol
5. Isopropyl alcohol
Summary of Contents for PM-7000
Page 1: ...PM 7000 Patient Monitor Service Manual ...
Page 2: ......
Page 58: ...Chapter 3 Product Specifications 46 line 2 5m ENF 350 ms Delay time 4s ...
Page 81: ......
Page 82: ...P N 7001 20 67407 1 0 ...