Chapter 3 Product Specifications
34
3.9 RESP Specifications
Measurement technique
Thoracic impedance
Lead
Optional: lead I and lead II; default lead II
Differential input
impedance
> 2.5M
Ω
Respiration impedance test
range
0.3 to 3
Ω
Excitation current
< 300µA
Baseline impedance range
200 to 2500
Ω
(using an ECG cable with 1k
Ω
resistance)
Bandwidth
0.2 to 2Hz (-3 dB)
Sweep speed
6.25 mm/s, 12.5 mm/s, 25 mm/s
RR
Measurement range
Adult:
Pediatric/neonate:
0 to 120 BrPM
0 to 150 BrPM
Resolution 1
BrPM
Precision
7 to 150 BrPM:
0 to 6 BrPM:
±2 BrPM or ±2%, whichever is greater.
Undefined.
Apnea alarm delay
10 to 40s