Chapter 3 Product Specifications
44
3.16 AG Specifications
Measurement technique
Infrared absorption
Measurement mode
Side stream
AG functions
Meets requirements of ISO9918, ISO11196, EN12598 and
ISO7767
Warm-up time
45 seconds (warming-up status)
10 minutes
(ready-to-measure status)
Adult/Pediatric
120, 150, 200 ml/minute (user-selectable)
Sampling flow
(sidestream)
Neonatal
70, 90, 120 ml/minute (user-selectable)
Gas type
CO
2
, N
2
O, O
2
(optional), Des, Iso, Enf, Sev, Hal
Measurement range
CO
2
:
N
2
O:
Des:
Sev:
Enf, Iso, Hal:
O
2
:
AwRR:
0 to 30%
0 to 105%
0 to 30%
0 to 30%
0 to 30%
0 to 105%
2 to 100 BrPM
Resolution
CO
2
:
AwRR:
1
mmHg
1 BrPM
Gas
Range (%REL)
Precision (%ABS)
0 to 1
±
0.1
1 to 5
±
0.2
5 to 7
±
0.3
7 to 10
±
0.5
CO2
> 10
Not specified
0 to 20
±
2
N2O
20 to 100
±
3
0 to 1
±
0.15
Precision
Des
1to 5
±
0.2
Summary of Contents for PM-7000
Page 1: ...PM 7000 Patient Monitor Service Manual ...
Page 2: ......
Page 58: ...Chapter 3 Product Specifications 46 line 2 5m ENF 350 ms Delay time 4s ...
Page 81: ......
Page 82: ...P N 7001 20 67407 1 0 ...