Other Functions
27
The Pulse Oximeter can also display technical error messages. In table
5-2, error messages as well as their causes and solutions are listed.
If the LCD screen can’t display anything, it may be damaged or error
occurs during system self-test. Please shut down the device (if can’t,
remove the batteries) and contact our Customer Service.
Table 5-2 Error indications
Error Message
Cause
Solution
Initiate Error
Failed
self-test
Shut down the device (if can’t, remove the
batteries) and contact us for service.
Please Release
the Button
Button
error
Check for jammed button. If problem
remains, contact us for service.
Pulse Not Found
Searching…
Pulse not
found
Check the patient and alert the doctor.