Measurement
22
Placement of a sensor on the same extremity with a blood
pressure cuff, arterial catheter, or intravascular line.
Loss of pulse signal can occur in the following situation:
The sensor is too tight;
There is excessive illumination from light sources such as a
surgical lamp, a bilirubin lamp, or sunlight;
A blood pressure cuff is inflated on the same extremity as the
one with a SpO
2
sensor attached;
The patient has hypotension, severe vasoconstriction, severe
anemia, or hypothermia;
There is arterial occlusion proximal to the sensor;
The patient is in cardiac arrest or in shock.