A - 5
Accuracy
Mechanical accuracy:
≤
±0.5%
Infusion accuracy:
≤
±1.8% (use Double-Dove and
B.Braun Original Perfusor Syringe, under standard
operating conditions, test in accordance with IEC60601-
2-24:2012)
Bolus accuracy: ±2% or 0.05ml, whichever is greater
(under standard operating conditions, test in accordance
with IEC60601-2-24:2012)
Set range of the infusion rate/
purge rage/bolus rate
Range of rate:
0.01 to 50ml/h(1ml syringe)
0.01 to 150ml/h (2/3ml syringe)
0.01 to 300ml/h (5/6ml syringe)
0.1 to 800ml/h (10/12ml syringe)
0.1 to 1200ml/h (20ml syringe)
0.1 to 1500ml/h (30/35ml syringe)
0.1 to 2300ml/h (50ml/60ml syringe and 60ml syringe)
Resolution:
0.01ml/h (0.01 to 99.99ml/h)
0.1ml/h (100.0 to 999.9ml/h)
1ml/h(1000 to 2300ml/h)
Occlusion pressure
15 levels selectable*: 50 mmHg, 150 mmHg, 225 mmHg,
300 mmHg, 375 mmHg, 450 mmHg, 525 mmHg, 600
mmHg, 675 mmHg, 750 mmHg, 825 mmHg, 900 mmHg,
975 mmHg, 1050 mmHg (not applicable for 50ml/60ml
syringe), and 1125 mmHg (not applicable for 50ml/60ml
syringe)
The maximum occlusion pressure is 1350 mmHg.
For the 2 ml syringe or syringes larger than 2ml, the selectable pressure ranges are as
follows:
• 0.01ml/h
≤
rate
<
0.1 ml/h: the selectable levels are 50 to 225 mmHg;
• 0.1ml/h
≤
rate
≤
100ml/h: the selectable levels are 50 to 1125 mmHg;
• 100ml/h
≤
rate
≤
2300ml/h: the selectable levels are150 to 1125 mmHg.
For the 1 ml syringe, the occlusion pressure is 975 mmHg.
Occlusion alarm tolerance
≤
±75 mmHg (for 50 mmHg level, 0.01ml/h
≤
rate
≤
100 ml/h)
≤
±20% or ±125 mmHg, whichever is greater (for 150 to
1125 mmHg levels, 0.1ml/h
≤
rate
≤
2300ml/h)
Maximum volume (under
single fault conditions)
≤
0.2ml
KVO rate
0.01 to 5.0ml/h
Minimum resolution: 0.01ml/h
Time set range
00:00:01 to 99:59:59
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...