4 - 6
NOTE
•
The delivered bolus volume will be added to the total infusion volume and
subtracted from the volume to be infused (VTBI).
•
The pump gives a beep every time a 0.5 ml bolus volume is infused.
4.6.1
Setting the Bolus Rate
To set the bolus rate, follow this procedure:
1.
Swipe the touchscreen from top down
→
select
Menu
→
select
General Option
.
2.
Set the
BolusRate
.
4.6.2
Automatic Bolus Infusion
To perform an automatic bolus infusion, follow this procedure:
1.
Select
from the main screen.
2.
Set the bolus volume in the popup dialog.
3.
Select
to start a bolus infusion.
The pump continues the infusion when the configured bolus volume has been infused.
If required, select
to stop the bolus infusion.
NOTE
•
If required, adjust the bolus rate in the BolusRate area during an automatic
bolus infusion.
4.6.3
Manual Bolus Infusion
To perform a manual bolus infusion, follow this procedure:
1.
Select
from the main screen.
2.
Set the bolus volume in the popup dialog.
3.
Press and hold
to deliver the required bolus.
4.
Release
when the desired bolus volume has been delivered or the bolus
volume limit is reached.
NOTE
•
The manual bolus volume limit is set in the User Maintenance menu. See
11.13 The Bolus Limit Settings
.
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...