NOTE
•
If the KVO rate is greater than the infusion rate, the pump will continue to
infuse at the set infusion rate.
•
The pump runs for 30 minutes at a KVO rate. At the completion of the KVO
infusion, the pump stops infusion, and gives a KVO Finish alarm.
•
The volume used during KVO infusion will be added to the total infusion
volume.
4.10
Unloading a Syringe
1.
In the main screen, select
to stop the infusion.
2.
Clamp the extension set.
3.
Disconnect the patient from the extension set.
4.
Open the door (1).
5.
Remove the extension set from the extension set holder (2).
6.
Pull down the syringe clamp(3), and remove the syringe from the pump (4).
7.
Proceed the next operation as needed:
◆
Continue the therapy: see
.
◆
Enter the standby mode: see
4.12 Entering the Standby Mode
◆
.
WARNING
•
Change the extension set as per the manufacturer’s instructions or the
hospital regulation.