4 - 7
4.6.4
Setting the Bolus Volume Unit
To set the bolus volume, follow this procedure:
1.
Swipe the touchscreen from top down
→
select
Menu
→
select
User
Maintenance
→
input the required password
→
select
.
2.
Select the
Bolus Volume Unit:
◆
ml
: the
BolusVTBI
unit is
ml
in each infusion mode.
◆
Dose
: in the
Dose Mode
or
Dose Time Mode
,
Drug Amt.
and
Volume
, or
Conc.
is set up, the
BolusVTBI
unit is the
Drug Amt.
unit or the
corresponding unit of the
Conc.
4.7
Changing the Infusion Parameters
You can modify rate, dose rate or drug name without stopping the infusion. This
function is called titration.
1.
Select the above parameters in the infusion running screen.
2.
Reconfigure the parameters in the popup dialogs.
To change other infusion parameters, follow this procedure:
1.
Press
to pause the infusion.
2.
Select the desired parameter area, and reconfigure parameters as per the
prescriber’s order.
3.
Select
OK
to confirm the changing.
4.
Press
again to resume the infusion.
4.8
Pausing the Infusion
Press
to temporarily stops a running infusion.
Press
again to restart the infusion after the infusion solution change.
4.9
Setting Keep Vein Open (KVO) Rate
At the end of infusion, the pump continues to infuse at a very low rate. KVO is used to
keep the patient’s vein open, to prevent back flow or vascular occlusion.
The default KVO rate is 0.5 ml/h. To edit the KVO rate, follow this procedure:
1.
Swipe the touchscreen from top down
→
select
Menu
→
select
General Option
.
2.
Set the
KVO Rate
. If
KVO Rate
is zero, the pump will not initiate a KVO infusion
when the preset volume is complete.
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...