12 - 6
NOTE
•
If the battery is not conditioned for a prolonged time, its charge indication
may not be accurate and you may wrongly evaluate the remaining battery
runtime.
•
Do not use the pump for infusion during battery conditioning.
•
Do not interrupt battery conditioning.
12.5
Checking Version Information
To view the system software version, brand library version, drug library upgrade time,
and Wi-Fi version, follow this procedure:
1.
Swipe the touchscreen from top down
→
select
Menu
→
select
User
Maintenance
→
input the required password
→
select
.
2.
Select
Version Information
.
12.6
Checking the History Record
The
History Record
menu shows the history of pump activities, including the infusions,
alarms, calibrations, maintenance configurations, and other operations.
To access the
History Record
menu, follow this procedure:
1.
Swipe the touchscreen from top down
→
select
Menu
→
select
System Options
.
2.
Select
History Record
.
NOTE
•
A total loss of power has no impact on the history records stored.
•
Alarms are saved as events and will remain if the equipment is powered
down. The time of equipment power down is also recorded as an event.
•
The pump stores up to 3500 events. When the capacity is reached, earlier
events will be overwritten by later ones.
12.7
Exporting the History Record
To export the history record, follow this procedure:
1.
Connect the USB drive to the USB connector.
2.
Swipe the touchscreen from top down
→
select
Menu
→
select
System Options
.
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...