A - 4
A.6
Wireless Network
A.7
Infusion Specifications
Standards IEEE
802.11a/b/g/n
Modulation mode
BPSK,QPSK, QAM
Operating frequency
2.412GHz to 2.484GHz
5.18GHz to 5.24GHz
5.745GHz to 5.825GHz
Data rate
IEEE 802.11a: 6 to 54 Mbps
IEEE 802.11b: 1 to 11 Mbps
IEEE 802.11g: 6 to 54 Mbps
IEEE 802.11n: 6.5 to 65 Mbps
Transfer power
< 20 dBm (CE requirement: detection mode – RMS)
< 30 dBm (FCC requirement: detection mode – PEAK)
Operating mode
Transmitting data through the wireless access point
(AP)
Data security
Standard: WPA-PSK and WPA2-PSK
Encryption: TKIP and AES
System capacity
Number of the pumps supported by a single AP:
≤
16
Data transmission delay
between the pump and the
CMS
Total data transmission delay time between the pump
and the CMS is
≤
8s
Interruption number and time
between the pump and the
CMS
Total interruption duration
≤
0.01* total
communication time
(Test within 24 hours, with 16 pumps, in which three
pumps are roaming for 30 times)
Delay time of network
disconnection alarm
≤
14 s
Compatible syringe sizes
1ml, 2ml, 3ml, 5ml/6ml, 10ml/12ml, 20ml, 30ml/35ml,
50ml/60ml (1ml is optional)
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...