11 - 3
11.5.3
The Central Station Setup
11.5.4
The Device Discover Settings
11.6
The Brand Management
S
NOTE
•
Up to 12 brands are available in this pump.
Menu Item
Default Setting
Function
Central Station IP
Address
0.0.0.0
Sets the central station IP address.
Menu Item
Default Setting
Function
Multicast TTL
1
Multicast helps device discovery between pumps
and between pumps and CMS. Devices in the
same multicast group can be mutually
discovered.
Multicast
Address
225.0.0.8
Menu Item
Default Setting
Function
Common Brand
/
Checks or unchecks the brand, and select
Confirm
.
The checked brand will be displayed in the brand
list.
Add Brand
/
Adds a brand by this procedure: input the brand
name
→
select a type (
Regular
, or
Light-
sensitive
)
→
select the syringe volume
→
select
Confirm
. The added brand is displayed in the
Common Brand
menu.
Delete Brand
/
Selects the undesired brand, and select
Confirm
to delete this brand.
Note:
The build-in brand is not allowed to be
deleted.
Modify Brand
/
Selects the brand that needs modifying, modify
this brand and select
.
Note:
The build-in brand is not allowed to be
modified.
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...