NOTE
•
The device needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided
below.
•
Other devices may affect this device even though they meet the
requirements of CISPR.
•
The EMISSIONS characteristics of this device make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this device
might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating
or re-orienting the device.
•
If the essential performance is lost or degraded, it may be necessary to take
mitigation measures, such as re-orienting or relocating the ME EQUIPMENT
or ME SYSTEM or shielding the location or stopping using the infusion pump
system and contact the service personnel.
Guidance and Declaration - Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment -
guidance
Conducted and radiated RF
EMISSIONS
CISPR 11
Group 1
The device uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic device.
Conducted and radiated RF
EMISSIONS
CISPR 11
Class A
The device is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic distortion
IEC 61000-3-2
Class A
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Voltage fluctuations and
flicker
IEC 61000-3-3
Complies
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...