Accumass BW500
PL-565
Par
am
et
e
rs
Function
Symbol
Description
PID mode
auto
manual
external alarm
the input contact status is sensed off
remote
communication
write
keypad / Dolphin Plus write (program) enabled
SmartLinx
®
/ remote device write (program)
enabled
Entry:
0 = off
f
1 = alternate display
2 = reset totalizer 1
3 = zero
4 = span
5 = print
6 = multispan selection*
7 = reserved
8 = reset batch
9 = PID freeze
10 = PID setpoint source
11 = PID mode
12 = external alarm
13 = remote communication write
*If the BW500 is programmed for multispan operation, the auxiliary input contact
state determines the multispan number (equivalent to P365). Input 1 is reserved for
multispan 1 and 2 selection. Input 2 is reserved for multispan 3 and 4 selection. Input
3 is reserved for multispan 5 to 8 selection.
multispan
selection
auxiliary input 1
auxiliary input 2
auxiliary input 3
1
2
3
4
5
6
7
8
If an attempt is made to select a multispan that has not been zero and span
calibrated, the selection request is ignored.
Notes:
•
When performing a remote span, it will first perform a zero, then it will ask you
to set up span test. Once loading is withing
±
2% of the design test weight, it will
perform the span.
•
For the print command to work, the BW500 must be in Run mode.
Summary of Contents for ACCUMASS BW500
Page 1: ...ACCUMASS BW500 Instruction Manual PL 565 January 2001 CCUMASS BW500 33455650 Rev 1 2...
Page 6: ...Page 6 Accumass BW500 PL 565...
Page 12: ...Page 12 Accumass BW500 PL 565 Specifications...
Page 28: ...Page 28 Accumass BW500 PL 565 Installation...
Page 88: ...Page 88 Accumass BW500 PL 565 Parameters...
Page 94: ...Page 94 Accumass BW500 PL 565 Operation...
Page 106: ...Page 106 Accumass BW500 PL 565 PID Control...
Page 110: ...Page 110 Accumass BW500 PL 565 Batching...
Page 112: ...Page 112 Accumass BW500 PL 565 Certification...