34
•
Adding jar model information to the database
To add new jar models and conductance reference values to the database, use BATTERY MANAGER
in the Main Menu. The BATTERY MANAGER available in BATTERY SETUP allows you to change
a reference value only for a particular data
fi
le. It does not save the change as a selection for new
test setups. You can also edit a model name or add a new model using BATTERY MANAGER in the
Main Menu.
Changing a reference value
1. Press the
ARROW
keys to highlight the BATTERY MANAGER in the Main Menu, and press
ENTER
.
2. Press the
ARROW
keys
to select a manufacturer, and press
ENTER
.
3. To change a reference value for a model in the database, press the
ARROW
keys
to select
the model name, and press
ENTER
.
4. In the
EDIT MFGR
screen, press the
DOWN ARROW
to select REFERENCE, and press
ENTER
.
NOTE:
If you change the model name and save it, the new name overwrites the original
model name in the database.
5. Press the
ARROW
keys
(or use the numeric keypad) to select a new reference value, and
press
ENTER
.
6. To save, press the
DOWN ARROW
to select SAVE AND CONTINUE, and press
ENTER
.
Adding a jar model
1. Press the
ARROW
keys
to highlight the BATTERY MANAGER in the Main Menu, and press
ENTER
.
2. Press the
ARROW
keys
to select a manufacturer, and press
ENTER
.
3. Press the
ARROW
keys
to select
one of the numbere
d USER DEFINED nam
es
, and press
ENTER
.
4. In the
MODEL ID
screen, your USER DEFINED selection is highlighted. Press
ENTER
to
change the name.
5. The cursor blinks to the right of the last space or character in the name. Display the charac-
ter, symbol, or number you want by rapidly pressing its key as many times as needed. If you
pause, the cursor moves to the right. To backspace one character, press the
BACK
key. To
enter a space between characters, press the
key twice. Press
ENTER
when
fi
nished.
6. In the
MODEL ID
screen, press the
DOWN ARROW
to select REFERENCE, and press
ENTER
.
7. Press the
ARROW
keys
(or use the numeric keypad) to select a new reference value.
8. To save, press the
DOWN ARROW
to select SAVE AND CONTINUE and press
ENTER
.
Chapter 5: Selecting test parameters
Celltron ULTRA