•
31
•
Setting the
LOW VOLTAGE
threshold
1. Press the
ARROW
(or use the numeric keypad) to highlight LOW VOLTAGE, and press
ENTER
.
2. Press the
ARROW
to scroll to the number of volts.
3. Press
ENTER
to accept.
Table 6: Low voltage threshold ranges and defaults
Range
Default
Range
Default
1.000 to 2.750 V for a 2 V jar
2.1 V
5.000 to 13.800 V for a 10 V jar
10.5 V
2.000 to 5.500 V for a 4 V jar
4.2 V
6.000 to 16.500 V for a 12 V jar
12.6 V
3.000 to 8.250 V for a 6 V jar
6.3 V
7.000 to 19.250 V for a 14 V jar
14.7 V
4.000 to 11.000 V for an 8 V jar 8.4 V
8.000 to 22.000 V for a 16 V jar
16.8 V
Selecting a test mode
(JARS ONLY)
The default is JARS ONLY.
1. Press the
ARROW
keys
to highlight JARS ONLY, JARS & 1 STRAP, or JARS & 2 STRAPS in
the
STRING INFO
list, and press
ENTER
.
2. Press the
ARROW
keys
(or use the numeric keypad) to check the box next to JARS ONLY,
JARS & 1 STRAP, or JARS & 2 STRAPS.
3. Press
ENTER
to accept.
Selecting the number of
POSTS PER JAR
The default is 2.
1. Press the
ARROW
keys
to highlight POSTS PER JAR, and press
ENTER
.
2. Press the
ARROW
keys
(or use the numeric keypad) to scroll to the number of posts.
3. Press
ENTER
to accept.
Chapter 5: Selecting test parameters
Celltron ULTRA