8.1
DAMAGED Firmware Recovery
If for some reason the firmware is damaged then the recovery is as follows:
Disconnect sensor from Unit
Select appropriate Virtual
Comport
Select Hex File with valid
Firmware.
Double Click the “Factory
Flash” label
Turn 24V power OFF
Push the MR340-1 Reset
button and hold.
Turn 24V Power ON
(keep button pressed)
Wait for 5 seconds and
observe the red and green
LED blink in alternating
fashion.
Release the Reset button.
Within 3 seconds time click
the Write Button on the
software.
The software will start
transmitting and writing the
new firmware. Wait until
completed.