Power-Up Sequence
Non-HOMING Linear Encoder
Step
Command Name Action
Notes
1
User actuator system is at initial reference position.
2
ZERO Input
Pulse ZERO Input
or send FC05
0x007 command
Controller will reset Position Counter to the
Preset Value.
3
System ready. Read Position Count and Speed via Modbus
Quadrature outputs can also be used independently.
4
An automatic Optical Calibration is performed whenever the sensor sweeps
through the active area at minimum speed of 0.2 meters/second.