4.4
M-Bus (if fitted)
After power-up, the unit defaults to the baud rate and primary address set in the M-Bus Menu (see
page 25). Both the baud rate and primary address may be changed later over the M-Bus network.
The secondary address is the unit serial number padded with two zeros.
Characters are configured as 8 data bits, 1 parity even bit and 1 stop bit, and the following bit rates
are supported: 300, 2400 and 9600 baud.
The U1000 MKII will only respond to M-Bus requests while operational, while the flow reading,
volume total, energy total, power or temperature screens are displayed.
The M-Bus module supports the following functions:
•
Acknowledge Function
•
Slave Select Function
•
Data Transfer Functions
•
Switch Baud Rate Function
•
Change Primary Address Function
4.4.1
Acknowledge Function
COMMAND:
ACK
DESCRIPTION:
Response from the slave denoting that a message was received from
the master.
DIRECTION:
SLAVE TO MASTER
FRAME TYPE:
ACK FRAME
NAME
CODE
ACKNOWLEDGE
0xE5