U1000MKII User Manual
Micronics
Page 18
Issue 3.1 July 19
4. Using the hose clips provided, clamp the Sensor Assembly (and adaptors, if used) to the pipe at
an angle of 45° to the top of the pipe. Experience has shown that the most consistently
accurate results are achieved when the unit is mounted at this angle (see page 46). This
minimises the effect of any flow turbulence resulting from entrained air along the top of the pipe
and sludge at the bottom.
Figure 19
Sensor Assembly clamped at 45°
2.6
Remove Sensor-Holding Screws
Release and remove the sensor-holding screws. The flow sensors are spring-loaded to ensure good
contact with the pipe surface.
NOTE: THE SENSOR-HOLDING SCREWS AND WASHERS SHOULD BE KEPT IN A SAFE
PLACE IN CASE IT IS NECESSARY TO RELOCATE THE UNIT (SEE PAGE 43).
Figure 20
Removing sensor-holding screws
45°
Air
Sludge