5
RELOCATING THE UNIT
If it is necessary to relocate the unit use the following procedure:
1. Disconnect the temperature sensors (U1000MkII-HM only) and MODBUS cable (if used).
2. Unfasten hose clips and remove the complete unit from the pipe.
3. Undo the screw at the end of the Sensor Assembly and gently lift the same end of the
Electronics Module as shown below.
4. The opposite end of the Electronics Module can now be released from the Sensor Assembly.
Figure 24
Separating Sensor Assembly and Electronics Module
5. Disconnect the two wires connecting the Sensor Assembly and Electronics Module.
6. Remove the original gel pads from the two sensors.
7. Push the sensor blocks into the Sensor Assembly so that the washers and locking screws can
be refitted.
8. Place replacement gel pads on the base of the sensors.
9. Follow the procedure for re-installing the unit on the pipe (see page 16).