2.3
Adjust Flow Sensor Separation
Using the separation code displayed by the Electronics Module (see page 14/15), take the Sensor
Assembly and adjust the flow sensor separation accordingly:
Figure 14
Loosen the flow sensor-holding screws (left); slide to correct position (right)
1. Undo the screws 2-3 turns, sufficiently to loosen the flow sensors and allow sideways
movement. DO NOT fully unfasten or remove the screws at this stage.
2. Slide the flow
sensors to the positions indicated on the display (e.g. “
D5
”).
3. With the flow sensors in the correct positions, tighten the sensor-holding screws so that the
sensors are secure.
2.4
Apply Gel Pads
1. Apply a gel pad centrally onto the bases of each of the two flow sensors.
2. Remove the covers from the gel pads.
3. Ensure there are no air bubbles between each pad and sensor base.
Figure 15
Applying the gel pads
2.5
Clamp Sensor Assembly to Pipe
The next step involves clamping the Sensor Assembly onto the pipe. Ensure that you have selected
a suitable location (see pages 6 and 46) and that the pipe is clean (see page 7). If you are installing
the unit on a pipe with an outside diameter less than 60mm use one or more of the adaptors
supplied with the unit.