Micronics
U1000MKII User Manual
Issue 3.1 July 19
Page 33
The following registers are available.
Register
Offset
Type
Typical
Contents
Meaning
Notes
n/a
Byte
0x01
Instrument Address
n/a
Byte
0x03
Instrument Command
n/a
Byte
0x40
Number of bytes to
read
0
Int-16
0x00
Device ID
0xAC
U1000MkII-FM/HM
0xac
1
Int-16
0x00
Status
0x0000 OK
Not[0x0000] Fault
0x00
2
Int-16
0x00
System Type
U1000MkII-HM only
0x04 Heating system
0x0C Chiller system
0x04
3
Int-16
0x00
Serial Identifier
0x01
4
Int-16
0x23
0x45
5
Int-16
0x60
0x00
6
IEEE754
float
0x40
Measured Velocity
Units in m/s
0x1f
7
0x67
0xd3
8
IEEE754
float
0x41
Measured Flow
Units in m
3
/hr for Metric
Units in US Gal/m for
Imperial
0x8c
9
0xd8
0xb0
10
IEEE754
float
0x42
Calculated Power
(U1000MkII-HM only)
Units in kW for Metric
Units in BTU/s for
Imperial
0x1c
11
0x2e
0x34
12
IEEE754
float
0x44
Calculated Energy
(U1000MkII-HM only)
Units in kWh for Metric
Units in BTU for
Imperial
0x93
13
0xc6
0xe8
Summary of Contents for U1000MKII-FM
Page 2: ......
Page 55: ...Micronics U1000MKII User Manual Issue 3 1 July 19 Page 51 7 DECLARATION OF CONFORMITY...
Page 56: ......