Page 91
Appendix| Factory Setting
optoNCDT 1420
A 2
Factory Setting
Password
„000“
Measurement averaging Median 9
Measuring rate
2 kHz
Output
Analog current
Measuring range
100 % FSO: I = 20 mA , digital 64877
RS422
921.6 kBaud
0 % FSO: I = 4 mA, digital 643
Trigger mode
No trigger
Peak selection
Highest peak
Language
German
Error handling
Error output, no measurement
LED
State
normal operation
Key
Select
Supply
voltage
red
flashes yellow approx. 8 Hz
flashes yellow approx. 1 Hz
yellow green
t 0
10 ... <15 s
t 1
t 3
t 2
10 s
t 4
Fig. 50 Flow chart to start a sensor with factory setting
t
0
:
power supply is on
t
1
... t
3
: both LEDs signalize the start sequence (red-yellow-green for 1 sec. each)
t
2
:
key is pressed during start sequence (t
1
... t
3
)
t
4
:
key is released while the LED
State
is flashing yellow
D
t = t
4
- t
2
;
D
t (key press period) must be at least 10 sec., max. 15 sec.
Summary of Contents for ILD 1420-10
Page 8: ...optoNCDT 1420...
Page 121: ......