Page 116
Appendix| Control Menu
optoNCDT 1420
A 4.2.3
Outputs
RS422
Baud rate
9.6 / 19.2 / 56.0 / 115.2 / 230.4 ... / 1000 kBps Transmission speed, binary data format
Output data
Distance / Exposure time / Intensity /
Sensor status / Measurement value counter /
non-inearized focus /
Timestamp / Video raw signal
Date to be transmitted are to be activated via the
checkbox.
Analog output
Standard scaling
Start of measuring range 4 mA,
at end of measuring range 20 mA
Two point scaling Start of range
Value
Always 2 points are taught which mark start and
end of new measuring range. With two point
scaling reversal of the output signal is possible.
End of range
Value
Digital output
Inactive
Regulates the switching performance of the
digital output (Error), see Chap.
.
Analog range: digital output switches when the
scaled analog range is exceeded.
Measuring range: The digital output is switched
when the peak is not (entirely) in evaluation
range (ROI).
Limit monitoring: digital output switches when
the limit is exceeded.
Analog range
Measuring range
NPN / PNP / PushPull / PushPullNeg
Limit monitoring
NPN / PNP / PushPull / PushPullNeg
Limit
Value
Hysteresis
Value
Minimum holding
period
Value
Data output
Web interface / Analog / RS422
Decides on the used interface for measurement
value output. A parallel physical measurement
value output via RS422 and analog is not pos-
sible. If web interface was selected, no measure-
ment values are output via RS422 or current
output.