Page 115
Appendix| Control Menu
optoNCDT 1420
Selection of peak
First peak /
Highest peak /
Last peak
Defines which signal is used
for the evaluation in the line
signal.
First peak: Nearest peak to
sensor.
Highest peak: standard, peak
with the highest intensity.
Last peak: widest peak to
sensor.
0
50
0
100
50
Range [%] 100
First
peak
Last
peak
Highest
peak
faraway
close
Sensor
Intensity
[%]
Error handling
Error output, no measurement value
The analog output supplies 3 mA instead of measurement value. The
RS422 interface outputs an error value.
Hold last value infinite
Analog output and RS422 interface stop at the last valid value.
Hold last value
1 ... 1024
Value
Measurement
value averaging
No averaging
Measurement values are not averaged.
Moving N values
2 / 4 / 8 ... 128
Value Indication of averaging mode. The averaging number N indicates the
number of consecutive measurement values to be averaged in the sen-
sor.
Recursive N
values
2 ... 32768
Value
Median N values
3 / 5 / 7 / 9
Value
Zeroing/
Mastering
Inactive
Normal measurement value resp. Zeroing/Mastering is undone.
Active
Value
Indication e.g. of thickness of a master part.
Value range 0 up to max. + 2 x measuring range
Reduction of
data
Value
Indicates the sensor which data is to be excluded from output, thus the
data amount to be transmitted is reduced.
Reduction is valid
for
RS422 / Analog
Interfaces to be used for undersampling are to be selected via the
checkbox.