Page 24
Installation
optoNCDT 1420
5.1.2.6
Surface Roughness
Laser-optical sensors detect the surface using an extremely small laser spot. They also track slight surface
unevenness. In contrast, a tactile, mechanical measurement, e.g. using a caliper, detects a much larger area
of the measurement object. In case of traversing measurements, surface roughnesses of 5 µm and more lead
to an apparent distance change.
Suitable parameters for the averaging number may improve the comparability of optical and mechanical
measurements.
h > 5 µm
Max.
Min.
Ceramic reference surface
Structured surface
Recommendation for parameter choice:
- The averaging number should be selected in such a way that a surface area the size of which is compa-
rable to those with mechanical measurements is averaged.