Set Sensor Parameter
optoNCDT 1420
7.4.4
Triggering
7.4.4.1
General
The optoNCDT 1420 measurement input and output is controllable through an external trigger signal or a
command. Triggering affects the analog and digital output.
The measurement value at the time of triggering is
output delayed, see Chap.
6.4
.
- Triggering does not influence the measuring rate resp. the timing so that between the trigger event (level
change) and the start of output always lie 3 1 cycle (Jitter).
- Micro-Epsilon recommends the abdication of data reduction, for example, by sub-sampling when the
triggering is used.
- The multifunctional input is used as external trigger input, see Chap.
5.4.6
.
- Factory setting: no triggering, the sensor starts data transmission right after start-up.
- Pulse duration of the “Trigger in“ signal must be at least 50
µs.
Triggering of measured value acquisition and output have the same time behavior.
Input trigger
/
Output
trigger
Level
A continuous measurement task is following as long as
the chosen level remains the same. Choice of level, see
Chap.
7.3
. Pulse duration must be at least a cycle time.
The following pause must be at least a cycle time..
Edge
infinite
Edge selection, see Chap.
7.3
. “0“ end trigger,
“1 ... 16382“ values per trigger, “16383“ endless trigger
manual Number Value
Software infinite
A software triggering is started by clicking the button
Re-
lease trigger
. “0“ end trigger
“1 ... 16382“ values per rigger, “16383“ endless trigger
manual Number Value
Inactive
No triggering
Grey shaded
fields require a
selection.
Value
Dark-bordered
fields require you
to specify a value.