A 1632
eMobility Analyser
Analyser operation
11
4
Analyser operation
The eMobility Analyser can operate autonomously or it can be remote controlled via Bluetooth
communication link with Metrel Instrument. Following test conditions can be set:
Keys, switches, LEDs
Relates to
Test condition
UINPUT On/Off = Off
INPUT
L/L1,L2,L3,N,PE
No voltage on the input of charging
cable, input is disconnected from mains
UINPUT On/Off = On
Error ON = Off
INPUT
L/L1,L2,L3,N,PE
Mains voltage NORMAL condition on
the input of charging cable
UINPUT On/Off = On
Error ON = On
ERRORS INPUT is set
INPUT
L/L1,L2,L3,N,PE
Mains voltage ERROR condition on the
input of charging cable (selected by
ERRORS switch position).
See chapter
PP STATE
OUTPUT
PP
EV simulation, charging cable current
rating
CP STATE
Error ON = Off
OUTPUT
CP
EV
simulation,
normal
operation
modes:
A, B, C no ventilation,
A, B, D ventilation required during
charging
Error ON = On
ERRORS OUTPUT is set
OUTPUT
CP
EV simulation, fault on CP simulated:
– Diode short,
– CP_PE short,
– PE open
Any combination
OUTPUT
L/L1,L2,L3,N,PE
Connection at the output of charging
cable / station. The condition depends
on the set state of the instrument.