79 - 79 021537/06/10
Value range: 0; 10...1000
Initial value: 60
Unit: Seconds
Command RP
<id>RP<para5> Read parameter data set
Access User mode
Description With this command a pre-stored data set can be loaded; see also Command SP. After
loading a restart is carried out automatically. The data set with the index ‘0’ contains the
parameters on delivery and cannot be overwritten.
The data sets ‘1’ and ‘2’ are available for use.
Value range 0..2
Command RS
<id>RS<para5> Restart ULTRASONIC (Reset)
Access: User mode
Description: With transmission of this command the ULTRASONIC watchdog is no longer operated.
This results in a cold start after approx. 2sec. The ULTRASONIC performs a complete
restart.
Parameter description:
1: The ULTRASONIC performs a cold start. It behaves as after connection of
the supply voltage.
Range of values 1
Initial value: No initial value
Command SC
<id>SC<para5> Minimum value of analog outputs (Start Current)
Access: User mode
Description: When using analog outputs the minimum value to be output can be selected. This does
not influence the terminal value of the outputs. The minimum value is 0% or 20% of the
terminal value. Depending on the parameter AN the selected minimum value is
converted to a current or voltage value. This provides for current and voltage outputs
with 0..20mA, 4..20mA, 0..10V, and 2..10V: see Command AN
See also: Command AA, Command AB
Parameter description:
0: Minimum value 0% of maximum value
1: Minimum value 20% of maximum value
Together with the parameter AN the analog outputs can be configured as follows:
Parameter
SC=0
Parameter
SC=1
Parameter AN=0
0..10V
2..10V
Parameter AN=1
0..20mA
4..20mA
Parameter AA= 0; AB = 0
Table 15: Configuration of analog outputs WV/RXD- and WD/RXD+ with parameters AN and SC