40
WIKA operating instructions, model CPC2000
11155094.04 11/2019 EN/DE/FR/ES
EN
11. Specifications
Base instrument
Functions
Menu languages
German, English, Spanish and French
Zero point correction
automatic (at definable time intervals)
manual (ZERO button)
Voltage supply
Supply voltage
DC 24 V, 1 A
Power consumption
24 VA
Battery type
Li-Ion
Battery life
approx. 8 h
Permissible ambient conditions
Operating temperature
10 ... 40 °C [50 ... 104 °F]
Storage temperature
-10 ... +70 °C [14 ... 158 °F]
Humidity
30 … 80 % r. h. (non-condensing)
Control parameters
Control steps
0 ... 50 % individually adjustable or 100 %
Control speed
approx. 5 s (dependent upon test volume)
Communication
Interface
RS-232 and USB
Response time
1 value/s
Certificates
Certificate
Calibration
3)
Standard: 3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
Recommended
recalibration interval
1 year (dependent on conditions of use)
3) Calibration in a horizontal position.
Approvals and certificates, see website
For further specifications see WIKA data sheet CT 27.51 and the order documentation.