23
WIKA operating instructions, model CPC2000
11155094.04 11/2019 EN/DE/FR/ES
EN
6. Commissioning, operation
3. Confirm with
[SELECT]
.
⇒
For the
AUTO
mode, the submenu points
RANGE
,
UNIT
and
STEPS
are not
relevant.
4. With the
[▲]
or
[▼]
buttons, select the menu item
SETTINGS
.
5. Confirm with
[SELECT]
.
⇒
The submenu item
ZERO
is not relevant.
6. In the submenu, select the item
AUTO-MODE
with the
[▲]
or
[▼]
buttons.
7. Confirm with
[SELECT]
.
The following items can be set in the operating mode AUTOMATIC-MODE:
Menu item
Meaning
RANGE
Pressure range to be measured
UNIT
Selection of the pressure unit (Pa, kPa, hPa, bar, mbar, psi, inH
2
O,
inHg, mmHg, Torr)
STEPS UP
The number of steps which will be taken to increase the pressure.
These step sizes are calculated by the controller itself.
STEPS DOWN
The number of steps which will be taken to decrease the pressure.
These step sizes are calculated by the controller itself.
CONFIGURATION 1
Further required settings for the
AUTO
mode can be set
In the submenu item
CONFIGURATION 1
, further settings can be defined for the
AUTO
mode.
Settings in the submenu item AUTO-MODE CONFIG 1
Menu item
Meaning
CYCLES
Here the number of cycles to be carried out in
AUTO
mode should be set.
T START
Delay time to vent the sensor
T STOP
Delay time after reaching the maximum defined pressure
T HOLD
Hold time until a new pressure value is reached
T PAUSE
Delay time between two cycles
AUTO ZERO
Automatic zero point correction at the completion of each cycle
AUTO-MODE CONFIG 1
CYCLES 001
T START 00005 s
T STOP 00005 s
T HOLD 00005 s
T PAUSE 00005 s
AUTO ZERO ON
-------------------------