21
WIKA operating instructions, model CPC2000
11155094.04 11/2019 EN/DE/FR/ES
EN
Function
Meaning
Leak test
By pressing the
[LEAK]
button, the control process will be deactivated
and the system sealed. It can be controlled via the display if the value
is stable. By pressing the
[LEAK]
button again, the mode will be
stopped and the instrument placed back in
Control
mode.
6.6 Operating mode Measure “MEAS”
The Measure mode is used to measure a differential or gauge pressure. The pump is
not active in this operating mode. The pressure is measured directly from the internal
reference sensor.
Setting
1. Press the
[MENU]
button.
2. Confirm the menu item
MODE
with
[SELECT]
and press the
[▲]
/
[▼]
buttons until
MEAS
is displayed.
3. Confirm with
[SELECT]
.
⇒
For the
MEAS
mode, the submenu items
RANGE
and
STEPS
are not relevant.
4. In the menu, select the item
UNIT
with the
[▲]
or
[▼]
buttons.
5. Select the pressure unit.
6. Confirm with
[SELECT]
.
7. With the
[MENU]
button, exit the menu.
SETTINGS submenu
1. With the
[▲]
or
[▼]
buttons, select the menu item
SETTINGS
.
2. Confirm with
[SELECT]
.
3. In the submenu, select the item
ZERO
.
4. Confirm with
[SELECT]
.
⇒
Under the point,
AUTO-ZERO
, the automatic zero point correction for the
MEAS
mode can be activated or deactivated.
6. Commissioning, operation