24
WIKA operating instructions, model CPC2000
11155094.04 11/2019 EN/DE/FR/ES
EN
6. Commissioning, operation
▶
To exit the submenu, press the
[MENU]
button.
Operation
On leaving the menu, the controller will revert to standby status. All information about the
AUTO
mode will be shown on the display.
The figure shows the flow chart for the
AUTO
mode. Based on the table, one will know
which cycles are running and the meaning of each.
Phase
Meaning
1
Wait for the start command with
[SELECT]
2
Delay time can be defined in the configuration
4
Duration of the system zero point correction
7
Hold time at the zero point
9
Time to control up to the next step
10
Hold time
13
Delay time at the maximum target pressure
1
Full scale
2
Pressure unit
3
Phase 1: it has been vented
(Pressure = 0)
4
Percentage display of the
current target pressure
5
Battery status
6
First of ten cycles
6
Operating mode
+10.000
VENT
mbar
+ 0% +0.000
Phase 1
Time 5 s
Press -0.007
Cycl.
1
/ 10
1
2
3
4
5
6
6
Start
1 2 4 5 7 9 10 9 10 9 10 9 10 13 15 16 15 16 15 16 15 16 18
Next cycle
End of cycle