7. Interface
Command
Answer function
:ats <1 ... 10000>
Start delay (phase 2)
1 ... 10000 --> time in seconds
Interface output
:o <0 | 1>
Output status information via interface
0 --> Off
1 --> On
Control mode
:pa <-110 ... 110>
Increase the target pressure specification by x percent
-110 ... +110 --> change in target pressure specification in %
:pd
Decrease target pressure specification by currently set step size in %
(Step DOWN)
.
:pr <-1100 ... 11000>
Adjust the current working and measuring ranges -1,100 ... +11,000
--> new measuring range in 0.01 % FS
:ps <-10 ... 110>
Percentage target pressure specification -10 ... +110 --> target
pressure specification in %
:pu
Increase target pressure specification by currently set step size in %
(Step UP)
.
:saz<0 | 1>
Set auto zero (in Measure and Control modes)
0 --> Off
1 --> On
:sbr<0 ... 1>
Set baud rate RS-232
0 --> 1200
1 --> 2400
2 --> 4800
3 --> 9600
4 --> 14400
5 --> 19200
6 --> 28800
7 --> 38400
8 --> 56000
9 --> 57600
:sbu<0 ... 1>
Set baud rate RS-232
0 --> 1200
1 --> 2400
2 --> 4800
3 --> 9600
4 --> 14400
5 --> 19200
6 --> 28800
7 --> 38400
8 --> 56000
9 --> 57600
:sci<n | u | r>
Select active interface
Set communication interface
n --> Interface off
u --> USB active
r --> RS-232 active
sdb <0 ... 100>
Display brightness
0 ... 100 --> brightness in %