33
WIKA operating instructions, model CPC2000
11155094.04 11/2019 EN/DE/FR/ES
EN
7. Interface
Command
Answer function
:spu <0 ... 9>
Pressure unit
0 --> Pa
1 --> hPa
2 --> kPa
3 --> mbar
4 --> bar
5 --> Torr
6 --> mmHg
7 --> inHg
8 --> psi
9 --> mmH
2
O
10 --> inH
2
O
:ssl <d | e>
System language
d = German
e = English
Menu settings
:smm <a | c | m >
Define the operating mode of the instrument
a --> start automatic mode
c --> start control mode
m --> start Measure mode
:ssw <1 ... 100>
Step size
1 ... 100 --> step size from operating the up and down buttons in %
:swm <a | c | m | z l v
| s>
Define the operating mode of the instrument
a --> start automatic mode
c --> start control mode
m --> start Measure mode
(only available in
CONTROL
and
MEASURE
modes)
z -->
ZERO
(only available in
CTRL
mode)
l -->
LEAK
test
l --> back to
CTRL
mode (like
[LEAK]
button)
v -->
VENT
(vent the entire system)
(only available in
AUTOMATIC
mode)
:szc <0.1>
Zero point adjustment before starting the control mode
0 --> Off
1 --> On
:szi <1 ... 60>
Zero point adjustment interval
1 ... 60 --> time in minutes
Additional commands
:sce <0.1>
Set communication interface echo
0 --> Echo Off
1 --> Echo On
On: Default. Sent command is returned, following the response, the
command will be concluded with “OK”.
Off: Only the pure response will be transmitted.