46
Menu functions
CAL 1
+0,5 K
CAL 3
-0,8 K
CAL 2
+2,6 K
180°C
40°C
120°C
0°C
factory calibration
Fig. 26
Schematic example of temperature adjustment
Example: Temperature deviation at 120 °C should be corrected.
1. Activate the adjustment setting. To do
so, press the activation key on the right
side of the
ADJUST
display. The display
is enlarged and the first calibration
temperature – in this case 40 °C –
automatically highlighted.
Cal1
°C
-0,2
K
Cal2
100.0
°C
+0,1
K
Cal3
180.0
°C
-0,2
K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
ADJUST
Temperature
Last adjustment 12.10.2012 12:00
Cal1
Cal2
Cal3
100.0
180.0
40.0
2. Press the confirmation key repeatedly,
until the calibration temperature CAL2 is
selected.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
180.0
40.0
100.0
3. With the turn control, set the calibration
temperature CAL2 to 120 °C.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
180.0
40.0
120.0
4. Save the setting by pressing the
confirmation key. The corresponding
calibration value is automatically
highlighted.
C
K
0,2
-
C
K
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
180.0
40.0
120.0
0,1
+