39
Menu functions
All other settings can be made accordingly. The settings possible are described in the
following sections.
If no new values are entered or confirmed for approx. 30 seconds, the appliance
automatically returns to the main menu and restores the former values.
7.3
Setup
7.3.1 Overview
In the
SETUP
display, you can set the following parameters:
►
the
IP address
and
Subnet mask
of the appliance's Ethernet interface (for connection to
a network)
►
the
Unit
of the temperature display (°C or °F, see page 40)
►
Alarm temp
: the temperature protection class according to DIN 12 880:2007-5 (TWW or
TWB, see pages 41 and 28)
►
the
Timer mode
(see page 42)
►
the heat output distribution (
Balance
, see 43)
7.3.2
IP address and subnet mask
If you want to operate one ore more appliances in a network, each appliance must have its
own unique IP address for identification. By default, each appliance is delivered with the IP
address 192.168.100.100.
LAN 1: 192.168.1.233
192.168.1.216
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
192.168.1