Menu functions
7.3.5
Timer mode
For universal ovens UN
PLUS
/UF
PLUS
and incubators IN
PLUS
/IF
PLUS
, you can choose if the timer (see
page 26) should run setpoint-dependent or not – this determines if the timer should not start
until a tolerance band of ±3 K around the set temperature is reached ( Fig. 23 , B) or if it should
start right after activation (A).
t
Timer
t
Timer
A
B
°C/°F
t
6 K
Fig. 23
Timer mode
A
Timer independent of setpoint: Timer starts right after activation
B
Timer setpoint-dependent: Timer does not start until tolerance band is reached
For sterilisers SN
PLUS
/SF
PLUS
, the timer is by default setpoint-dependent. To ensure that the
required temperature is maintained for a sufficient period of time, this setting cannot be
changed. If the temperature is no longer within the tolerance band, the sterilisation time
is, for reasons of safety, restarted as soon as the required temperature is reached again.
For universal ovens UN
PLUS
/UF
PLUS
and incubators IN
PLUS
/IF
PLUS
, the timer is in this case
interrupted and resumed as soon as the required temperature is reached again.
Setting
1. Activate the
SETUP
display and select
Timer mode
with the turn control.
Balance
Unit
IP address
-20 %
255.145.136.225
Subnetmask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer Mode
2. Accept the selection by pressing the
confirmation key. The adjustment options
are automatically highlighted.
Balance
Unit
IP address
-20 %
255.145.136.225
Subnetmask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer Mode
3. With the turn control, select the
desired setting – in this example Timer
independent of setpoint ( ).
Balance
Unit
IP address
-20 %
255.145.136.225
Subnetmask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer Mode