Technical data
Appliance size
30
55
75
110
160
260
450
750
Appliance width D
1
[mm]
585
585
585
745
745
824
1224
1224
Appliance height E
1
[mm]
707
787
947
867
1107
1186
1247
1726
Appliance depth G
1
(footprint) [mm]
434
514
514
584
584
684
784
784
Depth of door lock [mm]
56
Appliance depth F
1
(including door handle) [mm]
490
570
570
640
640
740
840
840
W
orking chamber width A
1
[mm]
400
400
400
560
560
640
1040
1040
W
orking chamber height B
1
[mm]
320
400
560
480
720
800
720
1200
W
orking chamber depth C
1
[mm]
250
330
330
400
400
500
600
600
Chamber volume [litres]
32
53
74
108
161
256
449
749
W
eight [kg]
37
45
51
68
72
91
125
163
Power [W]
IN/IF
230/115 V
800
1000
1250
1400
1600
1700
1700
1700
UN/UF/SN/SF
230 V
1600
2000
2500
2800
3200
3400
–
–
115 V
1100
1400
1700
2000
2200
2200
5800
2
7000
2
400 V
–
5800
7000
Current
consumption [A]
IN/IF
230 V
3.5
4.3
5.4
6.1
7.0
7.4
7.4
7.4
115 V
7.0
8.7
10.9
12.2
13.9
14.8
14.8
14.8
UN/UF/SN/SF
230 V
7.0
8.7
10.9
12.2
13.9
14.8
–
–
115 V
9.6
12.2
14.8
17.4
19.1
19.1
3 x 8.4
2
3 x 10.2
2
400 V
–
3 x 8.4
3 x 10.2
max. number of sliding shelves
346589
8
14
max. load per sliding shelve [kg]
30
max. load per appliance [kg]
60
80
120
175
210
300
300
300
Setting temperature
range
IN/IF
+20 to +80 °C
UN/UF
+20 to +300 °C
SN/SF
+20 to +250 °C
Adjustment precision
IN/IF
0.1 K
UN/UF/SN/SF
up to 100 °C: 0.1 K above 100 °C: 0.5 K
1
See Fig. 7
on page
15
2
3 x 230 V without zero