Operation and control
2. By turning the turn control, adjust the
desired lower alarm limit value, in the
example on the right 160 °C.
If no undertemperature protection limit is
required, set the lowest temperature.
ALARM
+/-
0.0
K
min
°C
auto
max
120
.0°C
160
0
.
3. Press the confirmation key to confirm.
The
max
display (overtemperature pro-
tection) is activated.
ALARM
+/-
0.0
K
min
°C
auto
max
°C
160
0
.
120
0
.
4. By turning the turn control, adjust the
desired upper alarm limit value, in the
example on the right 190 °C.
The monitoring temperature must be set
sufficiently high above the maximum set
temperature. We recommend 5 to 10 K.
ALARM
+/-
0.0
K
min
°C
auto
max
°C
160
0
.
190
0
.
5. Accept the upper alarm limit value by
pressing the confirmation key. The setting
of the automatic temperature monitor
(ASF) is automatically activated (
auto
).
ALARM
max
190
.0°C
+/-
0.0
K
min
160
.0°C
auto
6. With the turn control, select ON (
) or
OFF (
).
ALARM
max
190
.0°C
+/-
0.0
K
min
160
.0°C
auto
7. Press the confirmation key to confirm.
The ASF tolerance band setting is acti-
vated.
ALARM
K
min
°C
auto
max
190
.
0°C
160
0
.
+/-
0.0
8.
With the turn control, adjust the desired
tolerance band, e.g. 5.0 K.
We recommend a tolerance band of 5 to
10 K (1 to 3 K for incubators IN/IF).
ALARM
K
min
°C
auto
max
190
.
0°C
160
0
.
+/-
5.0
9. Press the confirmation key to confirm.
Temperature monitoring is now active.
ALARM
max
190
.0°C
+/-
5.0
K
min
160
.0°C
auto