26
Operation and control
5.4.5
Timer operation
In timer operation, you can adjust the time the appliance runs at the set value:
1. Press the activation key to the left of
the timer display. The timer display is
activated.
FA
TIMER
0
Ende 23.11. 09:00
00
h
0
m
2. Turn the turn control until the desired
duration is displayed – in this example 4
hours 30 minutes. The approximate end
time is shown beneath, in a smaller font.
TIMER
End
0
23.Nov 13:30
04
m
h
3
Up to a duration of 23 hours 59 minutes, the time is displayed in hh:mm (hours:minutes)
format. For 24 hours and more, the format dd:hh (days:hours) is used. The maximum
duration adjustable is 99 days 23 hours.
3. Press the confirmation key to confirm.
The display now shows the remaining
time in a large font and the approximate
end time in a smaller font beneath.
TIMER
30
m
04
h
End 23.Nov 13:30
4. Now, as described under 5.4.2 , set the individual values for temperature, air flap position
etc. which you want the appliance to operate at. For universal ovens UN
PLUS
/UF
PLUS
and
incubators IN
PLUS
/IF
PLUS
, the set values can be changed while the timer elapses. The changes
are effective immediately. For sterilisers SN
PLUS
/SF
PLUS
, parameters cannot be changed while
the timer elapses.
For universal ovens UN
PLUS
/UF
PLUS
and incubators IN
PLUS
/IF
PLUS
, you can choose if the timer
should run setpoint-dependent or not in the
Setup
– this determines if the timer should
not start until a tolerance band around the set temperature is reached or if it should start
right after activation (see page 42). If the timer runs setpoint-dependent, this is indicated by
the
symbol in the timer display.
When the timer has elapsed, the display shows 00:00. All functions
(heating etc.) are switched off. If a fan had been active, it will keep
on running for a short safety period.
To deactivate the timer, open the timer display by pressing the
activation key again and then turning the turn control to reduce
the timer setting until --:-- is displayed. Confirm with the confirma-
tion key.
TIMER
00
m
00
h
End 23.Nov 13:30
TIMER
--
m
--
h
End 23.Nov 13:30