15 Sterilisation chip card
15 Sterilisation chip card
As protection against contamination the HCP is equipped as standard with a STERICard. The STERICard
starts an automatic and pre-set sterilisation procedure.
To activate, insert the STERICard into the slit in the control field, select the Start symbol with the push/
turn control and the depressed SET key.
The auto-sterilisation sequence begins as soon as the sterilisation procedure is activated:
Ramp 1:
heating to 160°C
Ramp 2:
hold for 4 hours (after the sterilisation temperature has been reached)
Ramp 3:
cooling to 70°C
Ramp 4
wait 10 minutes
The automatic sterilisation programme can not be altered by the user.
After successful sterilisation the control displays
STERILISATION OK
. For documentation purposes the
sterilisation process can be reported on the STERICard and can be read using „Celsius 2007“.
Important note:
Important note:
On incubators the sterilisation programme does not serve for
On incubators the sterilisation programme does not serve for
sterilising the load but solely for sterilising the chamber. The incubator
sterilising the load but solely for sterilising the chamber. The incubator
is not a steriliser within the scope of the medical product regulations.
is not a steriliser within the scope of the medical product regulations.
Sterilisation
Process Control
Card
Sterilisation Parameter:
160 °C / 4:00 h
Unit-ID:
________________________