3.2 Standard equipment of HCP units
3.2 Standard equipment of HCP units
•
Electronic fuzzy-supported PID process controller with plus-package control and permanent power
proportioning and a time-saving autodiagnostic system for rapid fault finding (see section 22)
•
Recessing push/turn control for simple operation of oven
•
Language selection (English, French, Spanish, Italian, German)
•
Alphanumerical text display
•
Large-area all-round heating with additional heat conduction jacket
•
Capacitive humidity sensor
•
Active humidity control (fresh air via sterile filter) avoids condensation and ensures rapid
achievement of set humidity as well as short recovery times
•
Uniform atmosphere and temperature distribution through encapsulated and turbulence-free
ventilation system
•
Control of oven and documentation of actual values on MEMoryCard XL
•
Programme sequence control for up to 40 ramp segments
•
7 different ramp exit commands for advanced temperature applications
•
STERIcard for fully automatic sterilisation control during hot air sterilisation of complete cabinet
including sensors and fan disc
•
Integral weekly programmer with group function (e.g. all workdays)
•
Two separate PT100 temperature sensors Class A in 4-wire circuit for control and monitoring
•
Digital monitor controller for overtemperature, undertemperature, and as automatically setpoint
following monitor (ASF)
•
Mechanical temperature limiter (TB Class 1)
•
Monitor relay to switch off heating in case of fault
•
Visual alarm indication
•
Audible signal on over-/undertemperature, over-/under-humidity and open door
•
Internal documentation memory 1024kB as ring memory for all temperature and humidity values,
errors and settings with real-time and date. Documentation approx. 3 months at 1 minute storage
interval
•
Calibration possible without separate PC: 3-point calibration for temperature on the controller and
2-point calibration for humidity at 20% rH and 90% rH
•
Parallel printer interface (PCL3-compatible) for printing documentation data
•
Serial RS-232C interface for computer-supported temperature programmes and for reading the
internal report memory
•
MEMMERT software “Celsius 2007“ for control and documentation of temperature and relative
humidity via a PC
•
A pre-formatted blank MEMoryCard XL with 32 kB storage capacity, reprogrammable for up to 40
ramp segments and additionally 270 hours report memory at 1 minute intervals