HCP
page 29
40°C
34°C
ASF alarm
ASF activated
ASF activated
ASF-alarm
°C
40°C
34°C
ASF alarm
ASF activated
ASF activated
40°C
34°C
ASF-alarm
ASF activated
ASF aktiv
33°C
°C
27°C
ASF set to
+/-3°C in
+/-3°C in
+/-3°C in
SETUP
ASF set to
+/-3°C in
SETUP
SETUP
SETUP
Going outside tolerance margin = ASF alarm
Going outside tolerance margin = ASF alarm
Going outside the selected tolerance margin of the setpoint (in the example 37°C, +/-3°C), for example
through opening the oven door during operation, triggers the alarm.
Triggering the ASF alarm is indicated by flashing symbol
AUTO
and .
If the sounder is switched on in SETUP, the ASF alarm is additionally signalled by an interrupted tone. By
pressing the SET key the sounder can be switched off temporarily until the next occurrence of an alarm
event.
Again within tolerance margin = ASF alarm switched off
Again within tolerance margin = ASF alarm switched off
The automatic temperature monitor alarm is switched off automatically as soon as the selected tolerance
margin of the setpoint (in the example 37°C, +/-3°C) is reached again.